CareLit Fachartikel

Effect of Chemical and Radiochemical Impurities from Eluants on 99mTc-Labeling Efficiency

N.N. · Nuklearmedizin - NuclearMedicine · 1977 · Heft 06 · S. 241-246

Dokument
1025123
CareLit-ID
Jahr
1977
Publikation
PDF
nein
Metadaten
DOI
ja
zitierfähig

Bibliografische Angaben

Zeitschrift
Nuklearmedizin - NuclearMedicine
Autor:innen
N.N.
Ausgabe
Heft 06 / 1977
Jahrgang 16
Seiten
241-246
Erschienen: 1977-06-01 00:00:00
ISSN
0029-5566

Zusammenfassung

Summary Occasionally, radiopharmaceuticals prepared from the eluates of 99Mo-99m T c generators that have not been eluted for several days contain large amounts of free pertechnetate, as tested by radiochromatography and biologically by administration to patients. We find the most probable causes of poor 99mTc-labeling in these cases to be: a) the presence of a large concentration of 99TcO4 – in eluants; b) insufficient stannous ions available for the complete reduction of Tc due to spontaneous oxidation of Sn in the vial and also due to oxidation of stannous ions by the presence of larger than expected concentr…

Schlagworte

Nuklearmedizin - NuclearMedicine