CareLit Fachartikel

A novel medical suction device for improved visibility and accessibility of the uterotomy site during Cesarean section

BC Schmid, A Weghofer, GA Rezniczek, E Unger, P Husslein · Geburtshilfe und Frauenheilkunde · 2007 · Heft 06

Dokument
1076013
CareLit-ID
Jahr
2007
Publikation
PDF
nein
Metadaten
DOI
ja
zitierfähig

Bibliografische Angaben

Zeitschrift
Geburtshilfe und Frauenheilkunde
Autor:innen
BC Schmid, A Weghofer, GA Rezniczek, E Unger, P Husslein
Ausgabe
Heft 06 / 2007
Seiten
–
Erschienen: 2007-06-11 00:00:00
ISSN
0016-5751

Zusammenfassung

1. Fragestellung: Prototyping of a new suction device for uterotomy during Cesarean section. 2. Methodik: A novel fan-shaped suction device was tested in a sponge model simulating bleeding from a transverse uterine incision. Quantitative and qualitative differences in visibility of the operating field were measured. 3. Ergebnisse: Using the same suction force, 100% liquid was removed using the novel device compared to 51% using the standard Yankauer device. The novel device provided continuous visibility of 73% compared to maximal visibility of 43% using the Yankauer suction device. 4. Schlussfolgerung: The suct…

Schlagworte

Suction Blood Geburtshilfe und Frauenheilkunde