Effect of probe size and energy (M versus XL probe) on the detection of liver stiffness using transient elastography (Fibroscan)

E Durango, C Dietrich, W Vodovozov, A Kohlhaas, G Millonig, HK Seitz, S Mueller · Zeitschrift für Gastroenterologie · 2011 · Heft 08

Dokument
1079281
CareLit-ID
Jahr
2011
Publikation
PDF
nein
Metadaten
DOI
ja
zitierfähig

Bibliografische Angaben

Zeitschrift
Zeitschrift für Gastroenterologie
Autor:innen
E Durango, C Dietrich, W Vodovozov, A Kohlhaas, G Millonig, HK Seitz, S Mueller
Ausgabe
Heft 08 / 2011
Seiten
–
Erschienen: 2011-08-09 00:00:00
ISSN
0044-2771

Zusammenfassung

Background: Measurement of liver stiffness (LS) by transient elastography [TE/Fibroscan] is limited in patients with obesity and ascites. We here test and compare the recently developed more powerful XL probe versus the conventional M probe in obese and non-obese patients. Methods: LS was measured in 129 (BMI range 17.2–72.4) using both the M and XL probe. In addition, BMI, waist to hip ratio, transaminases and basic liver ultrasound parameters were obtained. Results: LS was not measurable in 6 patients (4.6%) by the XL probe as compared to 32 patients (24.8%) by the M probe. Of note, 46% of the patients not mea…