CareLit Fachartikel

Reduction of defective transducers by implementing regular technical quality assurance tests

S Wildner, C Kollmann · Ultraschall in der Medizin - European Journal of Ultrasound · 2016 · Heft 08

Dokument
1085215
CareLit-ID
Jahr
2016
Publikation
PDF
nein
Metadaten
DOI
ja
zitierfähig

Bibliografische Angaben

Zeitschrift
Ultraschall in der Medizin - European Journal of Ultrasound
Autor:innen
S Wildner, C Kollmann
Ausgabe
Heft 08 / 2016
Seiten
–
Erschienen: 2016-08-18 00:00:00
ISSN
0172-4614

Zusammenfassung

Purpose: Transducers are the most common error source for diagnostic ultrasound devices. Therefore regular technical quality assurance is intrinsic to be able to detect such faults at an early stage. With the implementation of a regular technical quality assurance concept for diagnostic ultrasound devices in the Austrian Mammography Screening Program the amount of defective transducers has been reduced significantly. Material and methods: Within two years 198 radiology departments including 237 ultrasound devices and 255 linear transducers were evaluated. Analysis of DICOM images from extensive initial tests, ad…

Schlagworte

Ultraschall in der Medizin - European Journal of Ultrasound