CareLit Fachartikel

A new Piezoelectric Power Source and First Animal Experiments

D. Neisius, J. Staudenraus, Y. Masur, M. Uder, Th. Gebhardt, M. Ziegler · Aktuelle Urologie · 1996 · Heft S 1 · S. 94-99

Dokument
1098585
CareLit-ID
Jahr
1996
Publikation
PDF
nein
Metadaten
DOI
ja
zitierfähig

Bibliografische Angaben

Zeitschrift
Aktuelle Urologie
Autor:innen
D. Neisius, J. Staudenraus, Y. Masur, M. Uder, Th. Gebhardt, M. Ziegler
Ausgabe
Heft S 1 / 1996
Jahrgang 27
Seiten
94-99
Erschienen: 1996-04-01 00:00:00
ISSN
0001-7868

Zusammenfassung

Summary: The disintegration efficacy of a new piezolectric power source has been measured in vitro using a standard plaster model stone. Shock wave pressure profiles were determined in the focal area by a PVDF-membrane hydrophone. Compared to the standard piezoelectric source the disintegration efficacy has been improved by about 50 %. 11 beagle dogs with 21 renal units were treated with 2000 shocks in a single session at the highest energy level by focusing the renal pelvis. Histopathological examinations were performed in 7 kidneys immediately after shock wave application, in 7 kidneys 6 weeks and in 7 kidneys…

Schlagworte

Extracorporeal Piezoelectric Lithotripsy (EPL) New power source Improved disintegration First animal experiments Significant renal damage immediately after EPL No permanent damage Aktuelle Urologie