Influence of Exposure Parameters and Iterative Reconstruction on MDCT-based Lung Densitometry – An ex vivo Phantom Study
P Leutz, MO Wielpütz, HU Kauczor, O Weinheimer, S Skornitzke, G Pahn, W Stiller, M Puderbach · Pneumologie · 2016 · Heft 07
Bibliografische Angaben
Zusammenfassung
Objectives: To evaluate the influence of exposure parameters and raw-data-based iterative reconstruction (IR) on lung densitometry on multidetector CT (MDCT) by dedicated in-house software. Material and Methods: 10 fresh porcine heart-lung-explants (4 simulated high, 6 simulated low attenuating chest wall) were prepared in an ex vivo chest phantom. MDCT scans were performed with 120 kV and 80 kV, each combined with 120, 60, 30 and 12 mAs, and reconstructed with both filtered back projection (FBP) and IR, resulting in 16 datasets per lung. Mean air density (AD), noise and lung density (LD) were measured by automa…