CareLit Fachartikel

In vitro quantification of tissue elasticity using three shear wave elastography platforms on liver fibrosis phantoms

A Mulabecirovic, A Batman Mjelle, OH Gilja, R Flesland Havre · Ultraschall in der Medizin - European Journal of Ultrasound · 2016 · Heft 08

Dokument
826265
CareLit-ID
Jahr
2016
Publikation
PDF
nein
Metadaten
DOI
ja
zitierfähig

Bibliografische Angaben

Zeitschrift
Ultraschall in der Medizin - European Journal of Ultrasound
Autor:innen
A Mulabecirovic, A Batman Mjelle, OH Gilja, R Flesland Havre
Ausgabe
Heft 08 / 2016
Seiten
–
Erschienen: 2016-08-18 00:00:00
ISSN
0172-4614

Zusammenfassung

Introduction: To assess and validate the reproducibility of quantitative elastography measurements, using shear-wave methods on four individual tissue-mimicking liver fibrosis phantoms with known Young's modulus. Methods: We used three different shear wave elastography platforms: GE Logiq E9 SWE, Philips iu22 XM ARFI and Samsung RS80A. Both linear (high frequency) and curvilinear (low-frequency) probes were applied. The objects were four individual tissue mimicking liver fibrosis phantoms with different Young's modulus within the range of soft biological tissue (2.7kPa, 11.5kPa, 24.8kPa, 46.3kPa). Two individual…

Schlagworte

Ultraschall in der Medizin - European Journal of Ultrasound