In vitro quantification of tissue elasticity using three shear wave elastography platforms on liver fibrosis phantoms
A Mulabecirovic, A Batman Mjelle, OH Gilja, R Flesland Havre · Ultraschall in der Medizin - European Journal of Ultrasound · 2016 · Heft 08
Bibliografische Angaben
Zusammenfassung
Introduction: To assess and validate the reproducibility of quantitative elastography measurements, using shear-wave methods on four individual tissue-mimicking liver fibrosis phantoms with known Young's modulus. Methods: We used three different shear wave elastography platforms: GE Logiq E9 SWE, Philips iu22 XM ARFI and Samsung RS80A. Both linear (high frequency) and curvilinear (low-frequency) probes were applied. The objects were four individual tissue mimicking liver fibrosis phantoms with different Young's modulus within the range of soft biological tissue (2.7kPa, 11.5kPa, 24.8kPa, 46.3kPa). Two individual…