Thyroid scintigraphy: establishing a clinically useful normal range for 99mTc pertechnetate uptake
Michael Grunert, Simone Agnes Schenke, Andrea Konrad, Christina Schütze, Stefan Förster, Burkhard Klemenz, Alexander R. Stahl · Nuklearmedizin - NuclearMedicine · 2024 · Heft 06 · S. 337-346
Bibliografische Angaben
Zusammenfassung
Abstract Purpose This study aims to establish a normal range for the thyroid uptake derived from 99mTc pertechnetate scans. In particular, variations of uptake with TSH stimulation and other factors such as urinary iodine concentration are taken into account and compared with the calculation of a raw uptake value. Methods Clinical multicentric (center A, B and C) prospective study on 125 consecutive healthy patients undergoing thyroid scans for thyroid nodules. Normal functional thyroid status was assured by normal TSH, normal thyroid size, no thyroid antibodies and no symptoms of thyroid functional disorders. C…