T Fuhrmann, R Kubale, W Krämer, J Ströder, A Buecker · Ultraschall in der Medizin - European Journal of Ultrasound · 2016 · Heft 08

Dokument
989071
CareLit-ID
Jahr
2016
Publikation
PDF
nein
Metadaten
DOI
ja
zitierfähig

Bibliografische Angaben

Zeitschrift
Ultraschall in der Medizin - European Journal of Ultrasound
Autor:innen
T Fuhrmann, R Kubale, W Krämer, J Ströder, A Buecker
Ausgabe
Heft 08 / 2016
Seiten
–
Erschienen: 2016-08-18 00:00:00
ISSN
0172-4614

Zusammenfassung

Purpose: To evaluate the correctness of shear wave measurements of different vendors and scanheads. Material and methods: For comparing different US machines and scan heads we used the 2nd generation elastography phantom of CIRS (039) with 4 different stiffness values from 3.5 kPa (1), 10 kPa (2), 25 kPa (3) and 45 kPa (4) simulating the range from normal liver to severe fibrosis. 3 US machines with up to three scan heads (4 MHz, 6 MHz and 9 MHz) were compared (Siemens S2000 and S3000 with 4C1, 6C1, and 9L4 scan head, Toshiba Applio 500 with 6C1 scan head). In every phantom 5 measurements are done in 2, 4, 6 and…